Troubleshooting
Hardware/PID testing:
| Problem | Diagnosis | Solution |
|---|---|---|
| Inconsistent, "squiggly" odor signal from PID | Likely something (probably part of the septum) caught in the needle, blocking air flow into the vial | Replace vial input needle |
| Odor signal from one line is significantly higher/lower than others | Variety of possible issues | Check if flow sensor is hitting desired setpoint. If problem persists, check for leaks. |
Electronics/GUI:
| Problem | Solution |
|---|---|
| Olfactometer not sending back flow values once turned on | If it is sending back slave addresses: If it's not sending back slave addresses: |
| Zero flow reported from flow sensor, even when proportional valve signal is at max (255) | Replace proportional valve |